Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Sanwa ED-570C: CAL ADJ /5500 DATE: 09-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.5 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 69 CONFIGURATION: Fluke 5500A ============================================================================= # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # This Procedure verified using a "generic" UUT. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP Procedure not formally tested with UUT. 1.001 DISP See procedure README file. 1.002 ASK- V 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Warm-up time: 30 min. - REQUIRED 1.004 DISP Temperature: 22 - 24 degC 1.004 DISP Humidity: <70% 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 DISP To prevent a shock hazard, remove all test leads 1.006 DISP from the instrument and turn the instrument OFF 1.006 DISP before disassembling the Model ED-570C. 1.007 HEAD DC VOLTAGE 1.008 DISP Connect the 5500A to the UUT as follows: 1.008 DISP [32] 5500A NORMAL HI to UUT V/OHM 1.008 DISP [32] 5500A NORMAL LO to UUT COM 1.009 RNG 320mV 1.010 5500 300mV S 2W 1.011 DISP Adjust trimmer TR3 for an indication of 300.0 +/-1. 1.012 5500 * S 1.013 RNG 3.2V 1.014 5500 3V S 2W 1.015 DISP Adjust trimmer TR1 for an indication of 3.000 +/-.01 1.016 5500 * S 1.017 HEAD AC VOLTAGE 1.018 RNG 3.2V AC 1.019 5500 3V 60H SI S 2W 1.020 DISP Adjust trimmer TR4 for an indication of 3.000 +/-.01 1.021 5500 * S 1.022 DISP Connect the 5500AS to the UUT as follows: 1.022 DISP [32] 5500A AUX HI to UUT 12A 1.022 DISP [32] 5500A AUX LO to UUT COM 1.023 RNG 12A 1.024 5500 10.00A S 2W 1.025 DISP Adjust trimmer TR2 for an indication of 10.00 +/-.1 1.026 END